Application of Radiation Protection of Patients in Medical Diagnostic Radiology (Case of Public Hospitals, Fianarantsoa District, Haute Matsiatra Region, Madagascar)

Received: 7 January 2021    Accepted: 25 January 2021    Published: 4 March 2021
Views:       Downloads:
Abstract

To strengthen patient radiation protection and control of medical exposure, in the Haute Matsiatra region, Fianarantsoa District, Madagascar, we have carried out a study to develop a protocol for dose assessment in diagnostic radiology in order to ensure good radiological practice. In this region, no studies have ever been done in their radiology department. The patient entrance dose (De) is one of the basic dosimetric quantities for measuring the dose delivered to the patients. Therefore, to assess the patient entrance dose (De), we have chosen two hospitals such as the Andrainjato University Hospital Center (CHUA) and the Tambavao University Hospital Center (CHUT). We have chosen four most requested radilogicals examinations (Skull, Thorax, abdomen and pelvis) and we have evaluated the entrance doses of patient and we have compared the result found to the Diagnostic Reference Levels (DRL) recommended by the IAEA for each examination. We found that, the average doses delivered to the patients during their radiographic examinations, were below the reference doses recommended by the IAEA. For this comparison carried out in these two hospitals, we can be confirmed that the values obtained can be useful for the application of regulations on the patient radiation protection and the control of medical exposure in Fianarantsoa.

Published in Nuclear Science (Volume 6, Issue 1)
DOI 10.11648/j.ns.20210601.11
Page(s) 1-4
Creative Commons

This is an Open Access article, distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution and reproduction in any medium or format, provided the original work is properly cited.

Copyright

Copyright © The Author(s), 2024. Published by Science Publishing Group

Keywords

Diagnostic Radiology, Patient Entrance Dose, Radiation Protection

References
[1] H. BEAUVAIS-MARCH, M. VALERO, A. BIAU, M. BOURGUIGNON; Niveaux de Référence Diagnostiques: spécificités de la démarche française en radiologie, 2003.
[2] K. E. M. Mohamadain, A. O. Osibote, A. C. P. Azevedo, L. A. R. Rosa, A. L. Lassance Cunha; Body Organ Doses in Conventional Radiology in Rio de Janeiro.
[3] E. K. Ofori, W. K. Antwi, L. Arthur and H. Duah; Comparison Of Patient Radiation Dose From Chest And Lumbar Spine X-Ray Examinations In 10 Hospitals In Ghana, 2012.
[4] Olivera Ciraj-Bjelac, Adnan Beganovic, Dario Faj, Vesna Gershan, Sonja Ivanovic, Ivan R. Videnovic, Madan M. Rehani; Radiation protection of patients in diagnostic radiology: Status of practice in fivenEastern-European countries, based on IAEA project; Science Direct; 2011.
[5] Kate Matthews, Patrick C. Brennan; The application of diagnostic reference levels: General principles and an Irish perspective; Science Direct; 2008.
[6] European Commission. European Guidelines on quality criteria for diagnostic radiographic images, EUR 16260 EN EN, Brussels, 1996.
[7] Harpen MD. A mathematical spread sheet application for production of entrance skin exposure. Med Phys, 1996; 23 (2) 241-242.
[8] International Atomic Energy Agency. Radiation Doses in diagnostic radiology and methods for dose reduction, IAEA-TECDOC-796, Vienna-Austria, 1995.
[9] International Atomic Energy Agency. International Basic Safety Standards for Protection against Ionizing Radiation and for the safety of Radiation Sources, IAEA Safety Series n° 115, Vienna-Austria, 1996.
[10] International Commission on Radiological Protection, Protection of the patients in diagnostic radiology. ICRP Publication 34, Pergamon Press, New York, 1982.
[11] International Commission on Radiological Protection, 1990 recommendations of ICRP, Publication 60, Pergamon Press, New York, 1991.
[12] International Commission on Radiological Protection, Radiological and Safety in Medicine, Publication 73, Pergamon Press, New York, 1996.
[13] Said Soudjay. Evaluation de la performance physique des dosimeters thermoluminescents pour la dosimétrie du patient en radiologie, Mémoire de DEA réalisé à Institut National des Sciences et Techniques Nucléaires (INSTN- MADAGASCAR), Faculté des Sciences, Université d’Antananarivo, 2007.
[14] Said Soudjay, Ramanandraibe Marie Jeanne, Ramanantsizehena Georgette Vololona, Raboanary Roland, Randriantseheno Hery Fanja, Rabesiranana Naivo, Randriamora Tiana Harimalala; Efficiency of Thermoluminescents Dosimeters for Patient Dose Assessment in Medical Radiology in Antananarivo, Science Publishing Group, 2019.
[15] Eric Kwasi Ofori, Diane Scutt, Matt Ward, William Kwadwo Antwi, Edmund Kwadwo Brakohiapa, Dzefi-Tettey, Klenam, Hewlett Kwaku Vincent; Estimation Of Patient Radiation Dose From Skull X-Ray Examinations In Ghana (Running Title: Patient Dose In Skull Examinations); Insuderc Academic Publishers; 2012.
[16] IAEA; Implementation of The International Code Of Practice On Dosimetry In Diagnostic Radiology; Vienna, 2011.
Cite This Article
  • APA Style

    Said Soudjay, Vololoniaina Bernardine, Ramanandraibe Marie Jeanne, Ramanantsizehena Georgette Vololona, Raboanary Roland, et al. (2021). Application of Radiation Protection of Patients in Medical Diagnostic Radiology (Case of Public Hospitals, Fianarantsoa District, Haute Matsiatra Region, Madagascar). Nuclear Science, 6(1), 1-4. https://doi.org/10.11648/j.ns.20210601.11

    Copy | Download

    ACS Style

    Said Soudjay; Vololoniaina Bernardine; Ramanandraibe Marie Jeanne; Ramanantsizehena Georgette Vololona; Raboanary Roland, et al. Application of Radiation Protection of Patients in Medical Diagnostic Radiology (Case of Public Hospitals, Fianarantsoa District, Haute Matsiatra Region, Madagascar). Nucl. Sci. 2021, 6(1), 1-4. doi: 10.11648/j.ns.20210601.11

    Copy | Download

    AMA Style

    Said Soudjay, Vololoniaina Bernardine, Ramanandraibe Marie Jeanne, Ramanantsizehena Georgette Vololona, Raboanary Roland, et al. Application of Radiation Protection of Patients in Medical Diagnostic Radiology (Case of Public Hospitals, Fianarantsoa District, Haute Matsiatra Region, Madagascar). Nucl Sci. 2021;6(1):1-4. doi: 10.11648/j.ns.20210601.11

    Copy | Download

  • @article{10.11648/j.ns.20210601.11,
      author = {Said Soudjay and Vololoniaina Bernardine and Ramanandraibe Marie Jeanne and Ramanantsizehena Georgette Vololona and Raboanary Roland and Randriamora Tiana Harimalala},
      title = {Application of Radiation Protection of Patients in Medical Diagnostic Radiology (Case of Public Hospitals, Fianarantsoa District, Haute Matsiatra Region, Madagascar)},
      journal = {Nuclear Science},
      volume = {6},
      number = {1},
      pages = {1-4},
      doi = {10.11648/j.ns.20210601.11},
      url = {https://doi.org/10.11648/j.ns.20210601.11},
      eprint = {https://article.sciencepublishinggroup.com/pdf/10.11648.j.ns.20210601.11},
      abstract = {To strengthen patient radiation protection and control of medical exposure, in the Haute Matsiatra region, Fianarantsoa District, Madagascar, we have carried out a study to develop a protocol for dose assessment in diagnostic radiology in order to ensure good radiological practice. In this region, no studies have ever been done in their radiology department. The patient entrance dose (De) is one of the basic dosimetric quantities for measuring the dose delivered to the patients. Therefore, to assess the patient entrance dose (De), we have chosen two hospitals such as the Andrainjato University Hospital Center (CHUA) and the Tambavao University Hospital Center (CHUT). We have chosen four most requested radilogicals examinations (Skull, Thorax, abdomen and pelvis) and we have evaluated the entrance doses of patient and we have compared the result found to the Diagnostic Reference Levels (DRL) recommended by the IAEA for each examination. We found that, the average doses delivered to the patients during their radiographic examinations, were below the reference doses recommended by the IAEA. For this comparison carried out in these two hospitals, we can be confirmed that the values obtained can be useful for the application of regulations on the patient radiation protection and the control of medical exposure in Fianarantsoa.},
     year = {2021}
    }
    

    Copy | Download

  • TY  - JOUR
    T1  - Application of Radiation Protection of Patients in Medical Diagnostic Radiology (Case of Public Hospitals, Fianarantsoa District, Haute Matsiatra Region, Madagascar)
    AU  - Said Soudjay
    AU  - Vololoniaina Bernardine
    AU  - Ramanandraibe Marie Jeanne
    AU  - Ramanantsizehena Georgette Vololona
    AU  - Raboanary Roland
    AU  - Randriamora Tiana Harimalala
    Y1  - 2021/03/04
    PY  - 2021
    N1  - https://doi.org/10.11648/j.ns.20210601.11
    DO  - 10.11648/j.ns.20210601.11
    T2  - Nuclear Science
    JF  - Nuclear Science
    JO  - Nuclear Science
    SP  - 1
    EP  - 4
    PB  - Science Publishing Group
    SN  - 2640-4346
    UR  - https://doi.org/10.11648/j.ns.20210601.11
    AB  - To strengthen patient radiation protection and control of medical exposure, in the Haute Matsiatra region, Fianarantsoa District, Madagascar, we have carried out a study to develop a protocol for dose assessment in diagnostic radiology in order to ensure good radiological practice. In this region, no studies have ever been done in their radiology department. The patient entrance dose (De) is one of the basic dosimetric quantities for measuring the dose delivered to the patients. Therefore, to assess the patient entrance dose (De), we have chosen two hospitals such as the Andrainjato University Hospital Center (CHUA) and the Tambavao University Hospital Center (CHUT). We have chosen four most requested radilogicals examinations (Skull, Thorax, abdomen and pelvis) and we have evaluated the entrance doses of patient and we have compared the result found to the Diagnostic Reference Levels (DRL) recommended by the IAEA for each examination. We found that, the average doses delivered to the patients during their radiographic examinations, were below the reference doses recommended by the IAEA. For this comparison carried out in these two hospitals, we can be confirmed that the values obtained can be useful for the application of regulations on the patient radiation protection and the control of medical exposure in Fianarantsoa.
    VL  - 6
    IS  - 1
    ER  - 

    Copy | Download

Author Information
  • Department of Physics, Faculty of Sciences and Technologies, University of Comoros, Moroni, Comoros

  • Department of Dosimetry and Radiation Protection, National Institute of Sciences and Nuclear Techniques, Antananarivo, Madagascar

  • Department of Dosimetry and Radiation Protection, National Institute of Sciences and Nuclear Techniques, Antananarivo, Madagascar

  • Department of Physics, Faculty of Sciences, University of Antananarivo, Antananarivo, Madagascar

  • Department of Physics, Faculty of Sciences, University of Antananarivo, Antananarivo, Madagascar

  • Department of Dosimetry and Radiation Protection, National Institute of Sciences and Nuclear Techniques, Antananarivo, Madagascar

  • Sections